ETSPL Study: Oscilla H210A Headphones
Equivalent threshold sound pressure levels (ETSPLs) for Oscilla H210A circumaural audiometric headphones
Published research · 2023
This peer-reviewed study was published in the International Journal of Audiology.
Bug M. U., Strüp M., Vollbort S., Fedtke T. · Published online 10 December 2023
What the study set out to do
The aim was to establish equivalent threshold sound pressure levels (ETSPLs) for the Oscilla H210A (Oscilla A30, A50, A60) circumaural audiometric headphones. ETSPLs are the reference values used to calibrate audiometers, so that a measured hearing threshold means the same thing from one device to the next.
How it was done
Thresholds were measured at 11 audiometric frequencies using the bracketing procedure described in ISO 8253-1. The measurements involved 25 otologically normal test subjects, using three sets of H210A headphones provided by Oscilla. No conflicts of interest were declared by the authors.
What was found
The study reports ETSPLs for the H210A measured on an ear simulator conforming to IEC 60318-1. It also includes a detailed measurement uncertainty budget, which shows how reliable each value is.
Why it matters
According to the authors, the values can be used to update ISO 389-8, the standard that defines reference ETSPLs for circumaural audiometric headphones.
Reference
Bug, M. U., Strüp, M., Vollbort, S., & Fedtke, T. (2023). Equivalent threshold sound pressure levels (ETSPLs) for Oscilla H210A circumaural audiometric headphones. International Journal of Audiology, 64(1), 89-93. Published online 10 December 2023. https://doi.org/10.1080/14992027.2023.2287394
Keywords: audiometry, ETSPL, equivalent threshold, bracketing procedure, Oscilla H210A, A30 Audiometer, A50 Audiometer, A60 Audiometer, headphones